Installing the Wash Container
Installing the Wash Container
The dispense tubing and dispense pump power cable must be connected.
To install the Wash Container:
1. Fill the Wash Container with the wash solution that is to be used.
Note: The Wash Container contains up to
4 liters.
2. Insert the Pick-Up Filter into the Wash Container and secure the
cap. Attach the Quick-Connector at the end of the pump tubing,
to the mating connector in the cap.
3. Route the dispense tubing through the upper slot at the front
corner of the instrument and through the routing clips at the front
of the instrument.
4. Connect the end of the dispense tubing to the wash tubing
connector on the wash head.
5. Thread the dispense tubing through the dispense valve.
(Figure 1).
Note: The tubing must be inserted fully into
the dispense valve opening to ensure proper
operation.
6. Connect the dispense pump power cable to the dispense pump
power connector at the rear of the instrument (Figure 3).