10 Chapter 1
Installation and Setup
Power Requirements
Installation and Setup
AC Power Cord
The analyzer is equipped with a three-wire power cord, in accordance
with international safety standards. This cable grounds the analyzer
cabinet when connected to an appropriate power line outlet. The cable
appropriate to the original shipping location is included with the
analyzer.
Various AC power cables are available that are unique to specific
geographic areas. You can order additional AC power cables for use in
different areas. AC Power Cords, on page 11 lists the available AC
power cables, illustrates the plug configurations, and identifies the
geographic area in which each cable is appropriate.
NOTE The front panel switch is a standby switch only; it is not a LINE switch
(power disconnecting device).
WARNING Install the product so that the detachable power cord is readily
identifiable and easily reached by the operator. The detachable
power cord is the product disconnecting device. It disconnects
the mains circuits from the mains supply before other parts of
the product. The front panel switch is only a standby switch
and is not a LINE switch. Alternatively, an externally installed
switch or circuit breaker (which is readily identifiable and is
easily reached by the operator) may be used as a disconnecting
device.
CAUTION Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can
cause product damage.
Table 1-1 Power Requirements
Description
Specification
1
Voltage, Frequency 100 to 120 Vrms, 50/60/400 Hz
220 to 240 Vrms, 50/60 Hz
Power Consumption, On Base Fully Loaded
< 260 W < 450 W
Power Consumption, Standby < 20 W
1. For greater detail, refer to the Specifications Guide.