Agilent Technologies E4440A Water Dispenser User Manual


 
Chapter 6 79
In Case of Difficulty
Before Calling Agilent Technologies
In Case of Difficulty
Perform an Alignment. Press System, Alignments, Align Now, All.
If the previously performed alignments did not resolve the problem,
press
System, Alignments, Restore Align Defaults. Then press System,
Alignments, Align All Now.
Is the analyzer displaying an error message? If so, refer to the
Instrument Messages and Functional Tests manual.
If the necessary equipment is available, perform the functional tests
in the Instrument Messages and Functional Tests document for your
analyzer.
If the 10 MHz output is not functioning, verify that it is turned on by
pressing
System, Reference, 10 MHz to select On.
TIP You can get automatic electronic notification of new firmware releases
and other product updates/information by subscribing to the Agilent
Technologies Test & Measurement E-Mail Notification Service for the
PSA or ESA Series at
http://wwwdb.tm.agilent.com/cgi-bin/notification/user.pl