Remote Commands
INPut Commands
OUTPUT 728;":INP:ATT?"
ENTER 728;A$
:INPut:LCMode
Syntax :INPut:LCMode <wsp> OFF|ON|0|1
Description This command sets the function of the
wavelength calibration. That is, whether the
wavelength calibration data is to be used to
reposition the filter to keep the attenuation
factor constant, or to alter the attenuation factor
with the filter kept in a fixed position.
Switch the mode on (using OFF or 0) to keep
the attenuation value fixed, and alter the filter
position. Switch the mode off (using ON or 1)
to keep the filter position fixed, and alter the
attenuation factor.
:INPut:LCMode?
Syntax :INPut:LCMode?
Description The query returns the current function of the
wavelength calibration.
0 indicates that the instrument is keeping the
attenuation value fixed, and altering the filter
position. 1 indicates the instrument is keeping
the filter position fixed, and altering the
attenuation factor.
Example OUTPUT 728;":INP:LCM ON"
OUTPUT 728;":INP:LCM?"
ENTER 728;A$
:INPut:OFFSet
Syntax :INPut:OFFSet <wsp>
<value>[DB]|MIN|DEF|MAX