127
Remote Commands
User Calibration Commands
The value that you send with this command, is
the attenuation for the next calibration point.
The wavelength of the calibration point is
updated automatically. The first piece of data is
for the start wavelength specified by the
:UCAL:START command. The default value
for the value is dB.
The value can be in the range 0.001dB to
99.999dB.
:UCALibration:VALue?
Syntax :UCALibration:VALue?
Description The query returns a value from the user
wavelength calibration data.
The value returned is the attenuation for the
next calibration point. The wavelength of the
calibration point is updated automatically. The
first piece of data is for the start wavelength as
returned by the :UCAL:START? query. The
values returned are in dB.
The error 204 indicates that there are no more
data points to be read.