Remote Commands
STATus Commands
Bit 7, which is 1 after the instrument has
repositioned the attenuator filter due to a
change in temperature.
Example OUTPUT 728;":STAT:OPER:COND?"
ENTER 728;A$
:STATus:OPERation:ENABle
Syntax :STATus:OPERation:ENABle <wsp>
<value>
Description This command sets the bits in the ENABle
register that enable the contents of the EVENt
register to affect the Status Byte (STB). Setting
a bit in this register to 1 enables the
corresponding bit in the EVENt register to
affect bit 7 of the Status Byte.
:STATus:OPERation:ENABle?
Syntax :STATus:OPERation:ENABle?
Description This query returns the current contents of the
OPERation:ENABle register.
Example OUTPUT 728;":STAT:OPER:ENAB
138"
OUTPUT 728;":STAT:OPER:ENAB?"
ENTER 728;A$
:STATus:OPERation[:EVENt]?
Syntax :STATus:OPERation[:EVENt]?
Description This query reads the contents of the
OPERation:EVENt register. Only three bits of
the event register are used (whether these bits
contain information depends on the transition
register configuration):