Making an Attenuation Sweep
The Manual Sweep
Figure 3-4 Running the Automatic Sweep
If there is something wrong with a parameter (if STEP is zero, for
example), this parameter is shown on the display for editing. Edit
the parameter, and press E
XEC again.
Repeating the Sweep
When the sweep is finished (SWEEP READY is shown at the
bottom of the display), you can press E
XEC to start it again.
Restarting the Sweep
To restart the sweep at any time while it is running, press EXEC.
3.3 The Manual Sweep
A manual sweep is one where stepping from one attenuation factor
to the next is done by the user.
Setting Up a Manual Sweep
There are three parameters for a manual sweep
• START is the attenuation factor at which the sweep begins.
• STOP is the attenuation factor that ends the sweep. If START
and STEP are such that the sweep does not end exactly at STOP,
then the sweep ends at the immediately previous value.