Programming the Attenuator
Some Notes about Programming and Syntax Diagram
Conventions
If more than 29 errors are put into the queue, the message -350
<Queue Overflow> is placed as the last message in the queue.
7.5 Some Notes about Programming and Syntax
Diagram Conventions
A program message is a message containing commands or queries
that you send to the attenuator. The following are a few points about
program messages:
You can use either upper-case or lower-case characters.
You can send several commands in a single message. Each
command must be separated from the next one by a semicolon
(;).
You end a program message with a line feed (LF) character, or
any character sent with End-Or-Identify (EOI).
You can use any valid number/unit combination.
Example 1500nm, 1.5
µm and 1.5e-6m are all equivalent.
If you do not specify a unit, then the default unit is assumed.
The default unit for the commands are given with command
description in the next chapter.
Short Form and Long Form
The instrument accepts messages in short or long forms. For
example, the message :INPUT:WAVELENGTH 1313 is in long
form, the short form of this message is :INP:WAV 1313.
In this manual the messages are written in a combination of upper
and lower case. Upper case characters are used for the short form of
the message. For example, the above command would be written
:INPut:WAVelength.