HP (Hewlett-Packard) 181A Water Dispenser User Manual


 
Chapter 2 Service
About the HP 53131A/132A Calibration Menu
2-12 Assembly-Level Service Guide
2
Equipment and Counter Setup: See Figure 2-1.
Figure 2-1. Calibration Setup for CAL: TI FINE?
Out of HP 8130A, into HP 59992A J06 Calibrator Signal:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low:
0.50 V
Input Mode: Normal
Timebase: Any external timebase you provide is ignored during
calibration.
Procedure: From the front-panel calibration menu, one keypress
invokes the calibration. You are prompted to press buttons on the
HP 59992A J06 Calibrator for each part of a four-part calibration.
Note that CAL: TI FINE? requires the completion of four steps in
order.
HP 53131A/132A
Counter
1234
In
AB
HP 59992A J06
Time Interval Calibrator
Start
Stop
HP 8130A
Pulse Generator
Output