HP (Hewlett-Packard) 181A Water Dispenser User Manual


 
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
1-58 Assembly-Level Service Guide
1
Test 4: Option 015/030/050/124 Channel 2
Frequency Sensitivity
This test verifies the frequency range and sensitivity of the optional
1.5-GHz (Option 015), 3-GHz (Option 030), 5.0-GHz (Option 050), or
12.4-GHz (Option 124) Channel 2 for the HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator (Not needed for Option 050/124)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (Not needed for Option 050/124)
HP 10503A BNC Cable (2)
HP 10100C 50
Feedthrough
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set the Counter as follows:
Freq & Ratio,
FREQUENCY 2