Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Assembly-Level Service Guide 1-37
1
Test 7: Peak Volts, Channels 1 and 2
(HP 53131A/132A Only)
This test verifies the Peak Volts accuracy specification of the HP 53131A
and HP 53132A Universal Counters.
Equipment
HP 3325B Synthesizer
HP 10100C 50
Ω
Feedthrough
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Other Meas key until
VOLT PEAKS 1
is displayed.
3 Set CHANNEL 1 input to:
50
Ω
/1M
Ω
, 50
Ω
X10 Attenuate, Off
100kHz Filter, Off
4 Set CHANNEL 2 input to:
50
Ω
/1M
Ω
, 50
Ω
X10 Attenuate, Off
100kHz Filter, Off