Gilson 402 Water Dispenser User Manual


 
Appendix C-4
Technical Data Appendix C
Hydraulics
Typical values (all syringes)
Syringe capacity Max. accuracy Max.error imprecision
%
+ A % RSD %
100 0.50 0.10
50 0.80 0.20
10 1.00 0.50
The same specifications are obtained in the
dispensing mode with liquid from the reservoir.
See OQ Document for procedure and calculations.
Syringe Volume Accuracy RSD
capacity (µL) pipetted (µL) % µL % µL
100 1 5.0 0.05 3.0 0.03
250 2.5 3.0 0.075 2.0 0.05
500 5 3.0 0.15 2.0 0.1
1000 10 3.0 0.3 1.0 0.1
5000 50 3.0 1.5 0.8 0.4
10000 100 3.0 3 0.5 0.5
25000 250 3.0 7.5 0.5 1.25
For 1 % syringe capacity, typical
values are given.
Junction Tubing
(between valve and Tee)
Material: FEP, with two PVDF union connectors.
Tubing dimensions: 2.0 mm i.d., 70 mm length.
Tubing dead volume: 220 µL.