Gilson 402 Water Dispenser User Manual


 
Appendix D-3
Error Messages Appendix D
Error Messages
Error Message Cause/Possible Action
Parameters incompatible! Refer to page 5-16.
Ratio impossible Dilution ratio calculated by software is not valid; check
parameters using formulae given in page 5-18 (DILUTE task)
and correct one or both of the other values.
Ratio missing Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Reservoir not allowed Liquids must be aspirated using the probe.
Right syr life exhausted! For the Dual syringe model: piston or right-hand syringe may
need servicing (refer to page 6-6).
Sample too small Enter a larger value for Sample Volume. Verify parameters in
the task.
Sample volume missing Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Syr capacity too small Check volumes used in the task correspond to the configured
syringe size(s).
Syringe Life Exhausted! For the Single syringe model: piston syringe may need
servicing (refer to page 6-6).
Total volume is missing! Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Total volume missing Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Total volume too small Enter a larger value for the Total Volume, or check the volumes
used in the task.
Valve life exhausted! Valve may need cleaning or servicing (refer to page 6-2).
xxx exists, overwrite? Select another file number, unless you want to overwrite file xxx.