5-28
Operation - Advanced Use 5
Key in the file number of the second Method to
link(8), press ENTER, press Next.
Key in the file number of the third Method to
link(9), press ENTER, press Next.
Key in the file number of the fourth Method to
link(8), press ENTER, press Next.
After a file is linked in this manner it can not be
deleted.
Press ESC to access the Save Screen, press Yes.
Typical Task Entry
Key to Press:
File # (44) ENTER
Edit
More
Link
File # (7) ENTER
Next
File # (8) ENTER
Etc until all file numbers
have been entered then:
ESC
Prefabricated Methods
Five prefabricated Methods, to run or to adapt, are
present in the 402’s memory as file numbers 90 to 94.
Although the tasks are the same, the parameters
vary according to whether you have a Single-
syringe or a Dual-syringe 402.
Configuration
Standard equipment with a 2 mL transfer tubing.
Left-hand (or only) syringe: 5 mL.
Right-hand syringe (if fitted): 500 µL.
Default values are used for parameters not listed.
Value for use with
File Number TASK/Parameter Dual-syringe Model Single-syringe Model
Method 90 Dilute Normal
Dilution Ratio 10 10
Sample Volume 200 1000
Rinse 300 1500
Method 91 Dilute Serially
Total Volume 3000 3000
Sample Volume 100 1000
Nb of Dilutions 3 3
Software