Performance Tests: Resident Self Tests
The two-digit codes that precede the submodule names are used in the Application Test Log
display, in its Category column, and in the Status window at the lower left of the window. See
“RST Test Log” on page 121.
Running an Interactive Test
The interactive tests check the operation of the system’s controls. They require you to make
selections and operate controls as they execute.
To run an interactive test:
1. Launch the RST application. (See “Launching the RST Application” on page 113.)
2. Log on using a valid service name and password.
3. Click the Te s t tab in the main RST window.
4. Click Interactive Test in the tree view, to show the available interactive tests.
5. Select the test you want to run.
NOTE Only one interactive test can run at a time.
6. Click Run. Testing begins.
7. Follow the instructions that the RST application displays.
8. To end a test, do the following:
To end a display test, press the THI key.
To end a keyboard test, right-click the test name, and select Stop from the pop-up
menu.