M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 59
Installation: Overview
5 Installation
Overview
This chapter includes procedures for installing and setting up the M2540A.
The following sections are in this chapter:
• “Assemble the System” on page 59
• “System Startup” on page 62
• “System Configuration” on page 64
• “Customer Training” on page 88
Assemble the
System
Complete the following steps to assemble the M2540A ultrasound system:
Power Cord and
Monitor
Installation
➤ Installing the Power Cord and Monitor
1. Attach the system power cord to the receptacle at the bottom rear of the system.
2. Loosen the screw that secures the cord retainer, rotate the retainer 180 degrees, and
tighten the screw.
3. Rotate the twivel base to its proper position. (The monitor cables pass through the rear of
the twivel base.)
4. Install the monitor on the twivel.
5. Connect the VGA cable the and monitor power cable to the monitor.
Peripherals
Installation
When installing peripherals, keep the following points in mind:
• No more than two peripherals can be mounted on the cart.
• When two peripherals are installed, the smaller one must be on top of the larger one.