Service Procedures: Transducer Connector Modules
Transducer
Connector
Modules
The screws that secure the transducer connector modules are behind the service panel on the
front of the lower enclosure.
Replacing Transducer Connector Modules
To replace a transducer connector module, see Figure 13-2 on page 183 to locate parts and
assemblies, and complete the following procedure:
1. Disconnect the system power cord from the main AC supply.
2. Remove the service panel at the right front of the lower enclosure by pressing down the two
latches at the panel’s top.
3. Remove the screws that secure the module bezel to the e-box, remove the bezel, and pull
the module out.
CAUTION Be careful when removing and installing the transducer connector modules, to prevent damage
to the distribution board’s connectors.
4. Reverse step 3 through step 1 to install a transducer connector module.
5. Perform all RST and system checks to verify performance.
Fan
The M2540A’s cooling fan is attached to the front of the power supply. Replacement fans are
supplied with the finger guard installed.
Replacing the Fan
To remove the fan, see Figure 13-2 on page 183 to locate parts and assemblies, and complete the
following procedure:
1. Disconnect the system power cord from the main AC supply.