M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 76
Installation: System Configuration
Figure 5-13 Field Service Options Window
3. Click Control Panel.
The Control Panel window opens.
4. Click Switch to Classic View.
The display changes to Windows Classic format.
5. Double-click Network Connections.
6. Double-click Local Area Connections.
7. Click the General tab.
8. Click Internet Protocol (TCP/IP).
9. Click Properties.
10. Enter the values supplied by the customer.
NOTE After opening the Control Panel window, you must reboot the system before it will
work as an ultrasound imaging device. This is true even if no settings are changed.