M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 79
Installation: System Configuration
Figure 5-15 DICOM Setup Window (Servers & Roles)
2. Select the correct server for each role from the drop-down lists.
3. To modify or verify attributes of storage or print servers, click Advanced next to the
server’s drop-down list.
4. Click Done.
➤ Configuring Autodelete
The following procedure configures how the system automatically deletes studies.
1. On the DICOM Setup window, click the This System tab.