Philips M2540 Food Saver User Manual


 
M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 128
Adjustments: Monitor Adjustment
Navigating the OSD Menus
The four buttons on the front of the monitor control all the OSD functions, as follows:
Brightness down (far-left button) applies the selected menu item.
Brightness up + Brightness down exits a menu without saving any changes.
Contrast down moves down one OSD Menu item or decreases the value of the selected
parameter.
Contrast up (far-right button) moves up to the previous OSD menu item or increases the
value of the selected parameter.
Adjusting VGA
Ver tical Size
For viewing BIOS settings mode, you must adjust the VGA vertical size. This adjustment does
not affect the normal display mode.
Adjust the VGA Vertical Size
1. Reboot the system.
2. As the system boots up, press the Patient key repeatedly, until the BIOS screen opens.
3. Disconnect the power cord from the rear of the monitor.
4. Press the two right-hand monitor buttons, and hold them in while reconnecting the power
cord.
5. Press the two center monitor buttons.
The OSD Main menu displays.