M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 88
Installation: Customer Training
Customer
Training
The customer needs some instruction in maintenance and use of the system:
• Instruct the customer how to back up and restore presets (see “Backing Up and Restoring
Presets” on page 143). Make a backup floppy disk of the customer’s presets and configura-
tions, and store it behind the lower enclosure’s left rear door. If the software is reloaded
later, the backup floppy disk lets you quickly restore settings and options.
• Explain the importance of letting the system shut down completely before disconnecting the
power cord.
• Demonstrate how to clean the air filter (“Cleaning the Air Filter” on page 131), and explain
the importance of the filter to system performance.
• Demonstrate how to remove and clean the trackball. (See “Cleaning the Trackball” on
page 132.)
• If the system is connected to a DICOM network, show the customer the job queue, and
explain that the queue is empty if all studies have transferred successfully.