M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 87
Installation: System Configuration
The driver is reinstalled with the correct configuration. This takes about 20 seconds.
7. Click Close to close the Setup window.
The new setting takes effect after the system is restarted.
Setting the Monitor Color Temperature
Cardiologists prefer a bluer tint to the monitor display than radiologists do. Pressing the two left
monitor buttons together toggles between the two color settings.
Setting 1 is appropriate for cardiology use.
Setting 2 is appropriate for non-cardiology use.
To set the monitor color temperature:
1. In normal ultrasound mode, press the two left monitor buttons at the same time.
2. A window displays the current color temperature setting.
3. To change the setting, press the two left buttons again, while the current setting is displayed.
Configuring for printers
The drivers for Sony UP-D895, Sony UP-D21MD, HP920, and HP1200 printers are loaded
automatically. These printers are ready to use when connected to a USB port.