M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 126
Performance Tests: Resident Self Tests
7. Right-click Service Log and select Save Log to File to save the entry.
Modifying an Entry in the Service Log
You can change the information in an existing service log entry. You must log on to the RST
application using a valid service name and password to modify an entry in the service log.
To modify a service log entry:
1. Launch the RST application. (See “Launching the RST Application” on page 113.)
2. Log on using a valid service name and password.
3. Click Service Log on the tree.
4. Double-click the log entry you want to modify. The Change Service Entry window displays.
5. Click in the entry field and modify the text.
6. Click OK.
7. Right-click Service Log and select Save Log to File to save the modification.