The following equipment is required to verify and troubleshoot the Agilent 6610xA Power Modules:
Table 1-1. Service Test Equipment
Equipment Required Critical Specifications Recommended Model Use
1
GPIB Controller Full GPIB capabilities HP Series 200/300 computer T
Oscilloscope Sensitivity: 1mV
Bandwidth Limit: 20MHz
Probe: 1:1 with RF tip
Agilent 54504A T
Rms Voltmeter True RMS Bandwidth: 20MHz
Sensitivity: 100
µ
V
Agilent 3400B T
Digital Voltmeter Resolution: 10nV @ 1V (8 1/2 digit) Agilent 3458A Accuracy: 20 ppm V,T
Variable Voltage
Transformer
Adjustable from- 13% to + 6% of input
range. 1KVA min.
T
CC RMS Noise
Resistor (R
M
)
Agilent 66101A,102A - 100A (.01
)
±
0.04%
2
Agilent 66103A,106A - 15A (.01
)
±
0.04%
3
Guildline 9230/100
Guildline 9230/15
V,T
Electronic Load Voltage and current range must exceed range
of module under test.
Power Range: 600W minimum
Agilent 6050A Load mainframe
and Agilent 60503A module
(240V) or Agilent 60507A module
(120V).
V,T
Load Resistor (R
L
)
Agilent 66101A- 0.1
±
5% 300W
Agilent 66102A-106A- 1.0
±
5% 300W
Ohmite C300KR10
Ohmite C300KIR0 T
MPS Mainframe Agilent 66000A V,T
MPS Keyboard
4
Agilent 66001A V,T
1 V = Verification; T = Troubleshooting.
2 Guildline Shunt Accuracy determined by Power Coefficient (
±
0.01% + (0.0004%/watt))
3 Guildline Shunt Accuracy determined by Power Coefficient (
±
0.01% + (0.002%/watt))
4 Use of the keyboard is recommended for displaying selftest error messages.