Agilent Technologies 6610XA Ventilation Hood User Manual


 
Verification and Performance Tests 19
Table 2-3. Performance Test Record for Model Agilent 66102A (20V, 7.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66102A Temperature___________________________Test Performed
By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0
±
8mV
DVM
±
3mV
1.6
µ
V
1.6
µ
V
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
20V
±
14mV
DVM
±
7mV
335
µ
V
335
µ
V
________
________
CV LOAD EFFECT 1mV
20
µ
V
________
CV SOURCE EFFECT 0.5mV
20
µ
V
________
CV NOISE (PARD)
Peak to Peak
Rms
7mV
3mV
0.872mV
500
µ
V
________
________
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0
±
3mA
Io
±
3mA
150
µ
A
150
µ
A
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
7.5A
±
5.3mA
Io
±
4.5mA
1mA
1mA
________
________
CC LOAD EFFECT 0.2mA
9
µ
A
________
CC SOURCE EFFECT 0.5mA
9µA
________
CC NOISE (RMS) 4mA 0.7mA ________