Agilent Technologies 6610XA Ventilation Hood User Manual


 
22 Verification and Performance Tests
Table 2-6. Performance Test Record for Model Agilent 66105A (120V, 1.25 A)
Pre Cal_____________Post Cal______________
Model Agilent 66105A Temperature___________________________Test Performed
By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0
±
54mV
DVM
±
32mV
2.1
µ
V
2.1
µ
V
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
120V
±
90mV
DVM
±
56mV
1.7mV
1.7mV
________
________
CV LOAD EFFECT 4mV
230
µ
V
________
CV SOURCE EFFECT 3mV
230
µ
V
________
CV NOISE (PARD)
Peak to Peak
Rms
25mV
18mV
1.0mV
1.5mV
________
________
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0
±
0.6mA
Io
±
0.6mA
15
µ
A
15
µ
A
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
1.25A
±
975
µ
A
Io
±
850
µ
A
157
µ
A
157
µ
A
________
________
CC LOAD EFFECT
50
µ
A1.4
µ
A
________
CC SOURCE EFFECT
50µA1.4µA
________
CC NOISE (RMS) 1mA 0.2mA ________