Agilent Technologies 6610XA Ventilation Hood User Manual


 
20 Verification and Performance Tests
Table 2-4. Performance Test Record for Model Agilent 66103A (35V, 4.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66103A Temperature___________________________Test Performed
By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0
±
13mV
DVM
±
8mV
1.7
µ
V
1.7
µ
V
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
35V
±
23.5mV
DVM
±
15mV
526
µ
V
526
µ
V
________
________
CV LOAD EFFECT 1mV
28
µ
V
________
CV SOURCE EFFECT 1mV
28
µ
V
________
CV NOISE (PARD)
Peak to Peak
Rms
10mV
5mV
0.90mV
500
µ
V
________
________
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0
±
2mA
Io
±
2mA
15
µ
A
15
µ
A
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
4.5A
±
3.4mA
Io
±
2.9mA
0.7mA
0.7mA
________
________
CC LOAD EFFECT 0.2mA
3
µ
A
________
CC SOURCE EFFECT 0.3mA
3µA
________
CC NOISE (RMS) 2mA
250µA
________