Agilent Technologies 6610XA Ventilation Hood User Manual


 
18 Verification and Performance Tests
Table 2-2. Performance Test Record for Model Agilent 66101A (8V, 16A)
Pre Cal_____________Post Cal______________
Model Agilent 66101A Temperature___________________________Test Performed
By_______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
Actual
CV PROGRAMMING @ 0 volts
Output Accuracy (DVM Reading)
Display and Readback Accuracy
0
±
3mV
DVM
±
2mV
1.5
µ
V
1.5
µ
V
________
________
CV PROGRAMMING @ full-scale
Output Accuracy (DVM Reading)
Display and Readback Accuracy
8V
±
5.4mV
DVM
±
3.6mV
88
µ
V
88
µ
V
________
________
CV LOAD EFFECT 1mV
1
µ
V
________
CV SOURCE EFFECT 0.5mV
1
µ
V
________
CV NOISE (PARD)
Peak to Peak
Rms
5mV
2mV
0.87mV
150
µ
V
________
________
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0
±
6mA
Io
±
6mA
150
µ
A
150
µ
A
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
16A
±
10.8mA
Io
±
9.2mA
2.1mA
2.1mA
________
________
CC LOAD EFFECT 0.5mA
15
µ
A
________
CC SOURCE EFFECT 0.75mA
15µA
________
CC NOISE (RMS) 8mA 0.9mA ________