Teledyne 6712FR Refrigerator User Manual


 
6712FR Refrigerated Sampler
Section 5 Extended Programming
5-36
5.19 Hardware Setup
The hardware setup screens make adjustments to the sampler’s
hardware-based features:
5.19.1 SDI-12 Sonde Setup The 6712 supports SDI-12 sondes. For more information on
setting up sondes, refer to Section 6.
5.19.2 Rain Gauge Setup With standard programming, the sampler does not record rain
gauge readings. However, with extended programming, recording
the rain gauge readings and the rain enable option can be turned
on or off. Enter the hardware setup and follow the screens to the
left in the margin.
When asked if you have a rain gauge connected, select YES to
record the rain gauge readings and add RAIN as a sampler
enable option. Select the appropriate tip for the rain gauge. The
sampler will then create a rain memory partition and update the
available sampler enable conditions.
Select NO to disable the readings and the sampler enable option.
When you select NO, the sampler removes the rain memory par-
tition, leaving the memory space available for other reading
types.
RUN “PROGRAM NAME
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
• Select Other Functions
MAINTENANCE
MANUAL FUNCTIONS
PROGRAMMING STYLE
HARDWARE
• Select Hardware
• SDI-12 Sonde Setup • Pager Numbers
• Rain Gauge Setup •Text Messaging
• I/O Pin Programming •Refrigerator Temperature
•Dialout Alarms •Analog Outputs
DO YOU HAVE A
RAIN GAUGE CONNECTED?
YES NO
Select YES.
RAIN GAUGE:
0.01 inch TIP
0.1 mm TIP
The rain gauges are set up to
tip on either .01 inch or .1 mm.
Select the desired tip.