Teledyne 6712FR Refrigerator User Manual


 
6712FR Refrigerated Sampler
Section 1 Introduction
1-12
Dry Weight: 145 lbs (66 kg) Refrigerator only, no controller or bottle configuration
160 lbs (73 kg) Includes refrigerator, controller, top cover, center section, tub,
pump tube, distributor shaft extension, distributor arm and nut, and discharge
tube for the 24 plastic bottle configuration.
Dry Weight With:
24 plastic bottles, rack, hold down, no caps 175 lbs(79 kg)
24 glass bottles, rack, expander ring, no caps 195 lbs(88 kg)
12 plastic bottles, rack, hold down, no caps 175 lbs(79 kg)
8 plastic bottles, rack, adaptor, no caps 170 lbs(77 kg)
8 glass bottles, rack, adaptor, no caps 180 lbs(82 kg)
2 plastic bottles, rack, no caps 175 lbs(79 kg)
2 glass bottles, rack, no caps 165 lbs(75 kg)
1-9.4 l plastic bottle and cap, no distributor arm or nut165 lbs(75 kg)
1-9.4 l glass bottle and cap, no distributor arm or nut 170 lbs(77 kg)
1-15.0 l plastic bottle and cap, no distributor arm or nut165 lbs(75
kg)
1-20.5 l plastic bottle and cap, no distributor arm or nut165 lbs(75
kg)
1-18.5 l glass bottle and cap, no distributor arm or nut170 lbs(77 kg)
Filled Bottle Weight: 24 plastic bottles 101 lbs(46 kg)
24 glass bottles 104 lbs(48 kg)
12 plastic bottles 89 lbs (41 kg)
12 glass bottles 95 lbs (43 kg)
8 plastic bottles 84 lbs (38 kg)
8 glass bottles 90 lbs (41 kg)
4 plastic bottles 98 lbs (45 kg)
4 glass bottles 104 lbs(48 kg)
1 plastic bottle 90 lbs (41 kg)
1 glass bottle 94 lbs (43 kg)
Average Thermal
Resistance Factor:
R-16
The average R-value for a complete FR refrigerator not running
Recovery time 5 minutes
Recovery time is the time for the internal air temperature to return to 39°F (4°C) after the door has been opened for
one minute. The empty refrigerator is running and set at 39°F (4°C). The door is opened at least 4 minutes after the
compressor has turned on. Ambient air temperature is 75°F (24°C).
Thermostat Set point range: 32° to 46°F (0° to 8°C)
Set point accuracy: ±1.8°F (±1°C) at 39°F (4°C)
[With an ambient air temperature of 75°F (24°C)]
Temperature Ranges Operational: -20° to 120°F (-29° to 49°C)
Storage: 0° to 140°F (-18° to 60°C)
Environment Rain and spray resistant, but not watertight. Do not install in locations where the
refrigerator’s lower compartment could become submerged.
Electrical
Power Requirements: 120 Volts, 60 Hz 230 Volts, 50 Hz
(optional)
Running current: 3.5 amp 1.75 amp
Starting current: 17 amp 8.5 amp
Stalled Compressor current:21 amp 10.5 amp
(maximum)
Table 1-3 Technical Specifications for the 6712FR Refrigerated Sampler (Continued)