Section 8 Maintenance
8-5
8.4.2 Pump Tube Alarm The sampler tracks the pump counts in both the forward and
reverse cycles with a resettable counter. When the counter
reaches the default count of 1,000,000 or the count you have
entered, the sampler displays a message, “WARNING!
REPLACE PUMP TUBE.” The message appears each time you
turn the sampler on or run a program until you reset the counter.
After replacing the pump tube (see page 8-9), reset the count to
zero so the sampler can begin tallying the pump counts for the
new tube. Replacing the pump tube does not reset the counter.
WARNING
Pump may actuate without warning. To avoid injury,
sampler must be off when pump housing cover is removed
for inspection or tubing replacement.
Experience may suggest a significantly different pump tube life.
You can change the alarm count to represent the pump tube life
for your application. The example below shows how to set the
alarm count and reset the pump counts.
Table 8-1 Setting the Clock and Calendar
1
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Select OTHER FUNCTIONS.
2
MAINTENANCE
MANUAL FUNCTIONS
PROGRAMMING STYLE
Select MAINTENANCE.
3
SET CLOCK
PUMP TUBE ALARM
INTERNAL BATTERY
DIAGNOSTICS
Select SET CLOCK.
4
ENTER TIME AND DATE:
HH:MM DD-MON-YY
14:00 22-JAN-01
Enter the time and date.
5
SET CLOCK
PUMP TUBE ALARM
INTERNAL BATTERY
DIAGNOSTICS
Press STOP to return to the main
menu.