Teledyne 6712FR Refrigerator User Manual


 
6712FR Refrigerated Sampler
Section 7 Remote Operation
7-3
Sampling reports You can retrieve the sampling reports at any time using the same
serial connection. Type “Q” to exit the menu. Then at the >
prompt, type “REPORT” and press Enter. The sampler controller
will send reports as configured.
Table 7-1 6712 Remote Menu Commands
Menu Command Description
0 or START This command will immediately start any program set to WAIT FOR PHONE CALL and cur-
rently waiting for the remote start command (see Section 4.6).
1 or RUN1 This command will load Program 1 as the sampler’s current Extended program, reset partition if
the storage interval has been changed, and run the program. RUN1 is valid when the sampler
is in standby, or when it is waiting for a start command. RUN1 will also run the current program
while in Standard programming.
2 or RUN2
3 or RUN3
4 or RUN4
These commands will load the selected program number as the sampler’s current Extended
program, reset partition data if the storage interval has changed, and run the program. They are
valid when the sampler is in standby, or when it is waiting for a start command. The commands
are available for Extended programming only.
5 or DISABLE This command is similar in functionality to a pin F low signal (a signal typically used to enable or
disable the sampler from a flow meter). At run time, the sampler starts with the sampler in the
enabled state. DISABLE is valid while a program is running. If successful, a display of REMOTE
ENABLE IS FALSE will be returned.
A Remote Disable is the highest priority disable. If the sampler is disabled with this command,
no condition other than menu command 6, ENABLE, will allow the program to become enabled
again.
6 or ENABLE This command is used to re-enable a disabled sampler. ENABLE is valid while a program is
running. If successful, a display of REMOTE ENABLE IS TRUE will be returned.
7 or TAKE_SAMPLE This command causes the sampler currently running a program to take a sample. The sample
will count as one of the program’s samples and is placed in the current bottle (Part A’s current
bottle when using two-part programming). TAKE_SAMPLE is valid while the program is run-
ning, after the start time, but not while in the MANUAL PAUSE screens. If successful, a
response similar to PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a
response similar to NO SAMPLE, PROGRAM WAITING START will be returned.
ST or STATUS This command causes the sampler to return its status information and current parameter data.
SD or
SCREEN_DUMP
This command causes the sampler to return the information currently displayed on the sam-
pler’s screen. The remote screen follows the sampler’s display as the information is updated.
This mode can be used to observe from a remote location while the sampler is programmed
on-site.
P or PAUSE This command causes a sampler currently running a program to enter the MANUAL PAUSE
menu. While you are connected in this mode, the sampler keypad becomes disabled, and your
computer’s keypad will be redirected to simulate the sampler keypad. The following keys will be
activated:
<Enter>, decimal point, and number keys.
<Esc>, S, or s = STOP.
<Backspace>, L or l, and U or u = Left/Up arrows.
R or r, and D or d = Right/Down arrows.