Teledyne 6712FR Refrigerator User Manual


 
6712FR Refrigerated Sampler
Section 5 Extended Programming
5-6
once enabled, stay enabled, no sample taken when the
sampler is enabled; first sample taken 20 minutes after
the sampler is enabled.
Part ‘B’ pause and resumes: None.
Start time: Delayed start of 1 minute.
730 Bubbler Module: Installed
Table 5-2 Storm Water Runoff Sampling
1
6712 SAMPLER
EXTENDED PROGRAMMING
For HELP at any
screen press ? key.
This screen will appear briefly.
2
RUN “EXTENDED 1”
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Select PROGRAM.
Note: If the start-up screen displayed “STANDARD PROGRAMMING”, type
6712.2 at this menu before selecting PROGRAM.
3
PROGRAM NAME:
“EXTENDED 1 “
CHANGE?
YES NO
Select YES.
Note: If the quick view screen shows up, press STOP then select OTHER
FUNCTIONS, SOFTWARE OPTIONS, QUICKVIEW/CHANGE, NORMAL.
Then press STOP and go back to step 2.
4
SELECT NEW PROGRAM
CHANGE PROGRAM NAME
Select CHANGE PROGRAM NAME.
5
NAME: “STORM
ABCDEFGHIJKLMNOPQRST
UVWXYZ-& 0123456789
BACK-UP DONE
Enter the word “STORM” for the program name.
6
SITE: DESCRIPTION
“FACTORY051“
CHANGE?
YES NO
Select YES.
7
SITE: “SITE 54
ABCDEFGHIJKLMNOPQRST
UVWXYZ-& 0123456789
BACK-UP DONE
Enter “SITE 54” for the site description.
8
SELECT UNITS FOR
LENGTH:
ft m
Select ft.
9
SELECT UNITS FOR
FLOW RATE
cfs gps gpm Mgd
lps m3s m3h m3d
Select gpm.
10
SELECT UNITS FOR
FLOW VOLUME
cf gal Mgal
m3 lit
Select gal.