Teledyne 6712FR Refrigerator User Manual


 
6712FR Refrigerated Sampler
Section 5 Extended Programming
5-42
Note
The internal analog output can not be installed in a controller
that has the internal voice modem also installed. Contact the
factory for information regarding external cellular modem
options.
Flowcharts for analog output screens appear in Appendix A.
5.20 Memory The 6712 Sampler contains 512 kilobytes of battery backed RAM
(Random Access Memory) that stores:
Five sampling programs
The sampling results from the most recent program
The rainfall data collected from rain gauges
The data from a refrigerator temperature sensor
The data collected from a 700 Series module
Parameter data collected from an SDI-12 Sonde
Calculated flow information from the current flow
module
The sampling results memory stores up to approximately 1004
events (the actual number is dependent on the number of SDI-12
parameters). The program run and start time information is
logged as four events. These four entries are held for the duration
of the program. The remaining 1000 are for logging sample
events, program events, and errors from the currently running
program. If the running program exceeds 1000 events, the data
will “roll over.” That is, the sampler will replace the oldest stored
event with the newest event. When using an SDI-12 sonde, the
sampler logs additional entries at each sample event.
Module, sonde, refrigerator temperature, and rain gauge
readings are stored in memory partitions. The storage capacity
depends on the reading types, the number of bytes per reading,
and the data storage interval.
The sampler clears the sampling results memory when the site
description changes. It is important to collect the sampling
results before changing the site description – the data can not be
retrieved afterwards. Keep in mind that the description may
change when selecting an extended program.
The partition data is cleared when the sampler automatically
creates and sizes the memory partitions. This occurs when you:
Select a new program with a different storage interval
Change the Data Storage Interval in the current
program
Change the programming mode
Install a new 700 Series Module
Change 700 Series Module mode of operation
Add or remove a rain gauge with the Hardware Setup