HP (Hewlett-Packard) HP 53131A Water Dispenser User Manual


 
Chapter 4 Command Reference
Introduction
Programming Guide 4-23
If the current measurement is Totalize or Voltage Peaks, Not a Number
9.91E37 is returned and error -221 is generated.
Query only.
The last calculated result remains valid until a new computation is made or a
relevant instrument state is modified.
:CALCulate3:FEED [:]CALCulate[1]
Sets or queries the data flow to be fed into the CALCulate3 block.
The string CALC is returned.
*RST: CALCulate[1]
:CALCulate3:LFILter Subtree
This subtree collects together the commands used to specify which measurements
will be used in computing statistics; out-of-limit measurements can be filtered out
of the statistics processing.
Not until :CALCulate3:LFILter:STATe is set to ON will any of the
:CALCulate3:LFILter settings be used.
:CALCulate3:LFILter:LOWer[:DATA] <numeric_value> [HZ | S
|DEG]
Sets or queries the statistics filter lower limit.
If limit filtering is enabled (:CALC3:LFIL:STAT ON), any measurements below
this value will not be combined into the statistics computation.
If math is enabled (:CALC:MATH:STATe ON), the limit value specified should
take into account that the filtering is on measurements that have been scaled and
offset.
-9.9999990000E+12 to -1.0000000000E-13, 0.0000000000, +1.0000000000E-
13 to +9.9999990000E+12.
11 digits
Comments
Query Response
Comments
NOTE
<numeric_value>
Range
<numeric_value>
Resolution