Chapter 4 Command Reference
Introduction
Programming Guide 4-121
:TRIGger Subsystem
This subsystem enables synchronization of instrument actions with specified
internal or external events.
:TRIGger:COUNt:AUTO <Boolean>
Sets or queries the control over the number of measurements made when
:INITiate[:IMMediate] is performed.
When :TRIG:COUN:AUTO is OFF, then :INIT[:IMM] initiates a single
measurement.
When :TRIG:COUN:AUTO is ON and [:SENS]:FUNC[:ON] is not Totalize or
Voltage Peak function, then:
when statistics are enabled (:CALC3:AVER[:STAT] is ON), :INIT[:IMM]
will initiate a complete block of N (:CALC3:AVER:COUN) valid
measurements.
when statistics are enabled (:CALC3:AVER[:STAT] is ON),
and limit-filtering is enabled (:CALC3:LFIL:STAT is ON), :INIT[:IMM] will
initiate a complete block of N (:CALC3:AVER:COUN) valid, in-limit
measurements.
when statistics are disabled (:CALC3:AVER[:STAT] is OFF), :INIT[:IMM]
initiates a single measurement.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
The :TRIG:COUN:AUTO setting has no affect on the operation of
:INIT:CONT ON, which always behaves as if :TRIG:COUN is set to 1.