HP (Hewlett-Packard) HP 53131A Water Dispenser User Manual


 
Chapter 4 Command Reference
Introduction
Programming Guide 4-101
The [:SENS]:TINT:ARM:EST ART[:LAYer[1]]:ECO command has no affect
on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse
Width.
[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:SOURce
IMMediate | TIMer | INTernal[1]
Sets or queries the start arm delay for Time Interval measurements.
A sequence of ASCII-encoded bytes: IMM or TIM or INT[1]
*RST: IMMediate
The [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR command has no
affect on the following measurements: Risetime, Falltime, Duty Cycle, and
Pulse Width.
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:TIMer
<numeric_value> [S]
Sets or queries the time used to delay the start arm for Time Interval
measurements.
100E-9 to 0.999,999,9 seconds
100E-9 seconds
Numeric data transferred as ASCII bytes in <NR1> format.
*RST: 100E-9 seconds
Only applies when [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR TIM is
selected.
The [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:TIM command has no affect
on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse
Width.
Gate & ExtArm
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
<numeric_value>
Resolution
Query Response
Comments
Related
Front-Panel
Keys