Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-64 Programming Guide
Descriptions of the Measurement Functions<function> (Cont.)
:MEASure[:SCALar][:VOLTage]:FREQuency? (Cont.)
default: value which indicates 4 digits of resolution for the
specified <expected_value>
Ch3 <expected_value>
range: 100 MHz to 3.00 GHz
resolution: <expected_value> should be within 10% of input
frequency for optimum arming configuration
default: 500 MHz
Ch3 <resolution>
description: value indicates decade corresponding to least significant
digit of the result
range: 1E-7 to1E7 Hz
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
resolution: <resolution> should use a mantissa of 1.0 and be an
even power of 10
default: value which indicates 4 digits of resolution for the
specified <expected_value>
<source_list>
range: (@1) | (@2) | (@3)
default: (@1)