HP (Hewlett-Packard) HP 53131A Water Dispenser User Manual


 
Chapter 4 Command Reference
*OPC
(Operation Complete Command)
4-132 Programming Guide
*OPC
(Operation Complete Command)
*OPC
This event command enables the OPC bit (bit 0) in the Standard Event Status
Register to be set upon the transition of the measurement cycle from measuring to
idle. (Note For HP 53131A/132A with serial number prefix 3646 and above
If the measurement cycle is already idle, this command will immediately set the
OPC bit.) See the section titled “Standard Event Status Register Group,page 3-
24, in Chapter 3 of this guide for a detailed description of the Standard Event
Status Registers Operation Complete bit.
This event command is “disabledby *CLS, *RST, Device Clear
(page 4-31), power-on, or upon the transition of the measurement cycle from
measuring to idle.
This event command has no query form.
See the section titled “Using the *OPC Command to Assert SRQ,
page 3-48, in Chapter 3 for an example using this command.