Agilent Technologies Model A.08.xx Water Dispenser User Manual


 
Chapter 5 213
Language Reference
CALCulate:LLINe Subsection
Merge Additional Values into the Existing Limit Line
:CALCulate:LLINe[1]|2:DATA:MERGe
<x-axis>,<ampl>,<connected>{,<x-axis>,<ampl>,<connected>}
Adds the points with the specified values to the current limit line, allowing you to
merge limit line data. Up to two amplitude values are allowed for each x value. If
too much data is merged, as many points as possible are merged into the existing
limit and then an error is reported. Up to 200 points total may be defined for each
limit.
<x-axis> can be frequency or time values as specified by
:CALCulate:LLINe:CONTrol:DOMain. Frequencies are always in Hz.
Time is always in seconds. No unit is allowed in this parameter.
<ampl> amplitude values are in the current Y-axis units. No unit is allowed in
this parameter.
<connected> connected values are either 0 or 1. A 1 means this point should be
connected to the previously defined point to define the limit line. A 0 means
that it is a point of discontinuity and is not connected to the preceding point.
The connected value is ignored for the first point.
Range: <x-axis> 30 Gs to +30 Gs for time limits
<x-axis> 30 GHz to +350 GHz for frequency limits
<ampl> 120 dBm to +100 dBm
<connected> 0 or 1
Front Panel
Access:
Display, Limits, X Axis Units Freq Time
Delete Limit Line
:CALCulate:LLINe[1]|2:DELete
Deletes the selected limit line.
Display the Limit Line
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
Controls the display of the current limit line.
Factory Preset
and *RST: Off
Front Panel
Access:
Display, Limits, Limit 1|2, Limit On Off