224 Chapter 5
Language Reference
CALCulate:MARKer Subsection
Factory Preset
and *RST: Off
Remarks: When a measurement under the front panel MEASURE key is
started, this command is turned off. If this command is turned
on when any of the MEASURE key measurements are in
progress, that measurement will be stopped.
Front Panel
Access:
FREQUENCY/Channel, Signal Track On Off
Marker X Value
:CALCulate:MARKer[1]|2|3|4:X <param>
:CALCulate:MARKer[1]|2|3|4:X?
Position the designated marker on its assigned trace at the specified trace X value.
The value is in the X-axis units (which is often frequency or time).
The query returns the current X value of the designated marker.
Default Unit: Matches the units of the trace on which the marker is positioned
Front Panel
Access:
Marker
Span Markers Center Frequency X Value
:CALCulate:MARKer[1]|2|3|4:X:CENTer <param>
:CALCulate:MARKer[1]|2|3|4:X:CENTer?
Position the center frequency, of the designated span-type marker pair, at the
specified trace X value. The value is in the X-axis units (which is often frequency
or time) Use :CALCulate:MARKer:MODE SPAN to select span markers.
The query returns the current X value center frequency of the designated markers.
Range: Matches the units of the trace on which the markers are
positioned
Front Panel
Access:
Marker, <active marker>, Span Pair
Marker X Position
:CALCulate:MARKer[1]|2|3|4:X:POSition <integer>
:CALCulate:MARKer[1]|2|3|4:X:POSition?
Position the designated marker on its assigned trace at the specified X position.
The query returns the current X position for the designated marker.
Range: Refer to the [:SENSe]:SWEep:POINts command.