236 Chapter 5
Language Reference
DISPlay Subsystem
Factory Preset
and *RST: The factory default is On. This parameter is persistent, which
means that it retains the setting previously selected, even
through a power cycle.
Front Panel
Access:
System, Time/Date, Time/Date On Off
Display Annotation Title Data
:DISPlay:ANNotation:TITLe:DATA <string>
:DISPlay:ANNotation:TITLe:DATA?
Enters the text that will be displayed in the user title area of the display.
Front Panel
Access:
Display, Title
Display, Title, Change Title
Display, Title, Clear Title
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
Turns the display on or off. Having the display turned off may increase repetitive
measurement rate.
The following key presses will turn display enable back on:
1. If in local, press any key
2. If in remote, press the local (system) key
3. If in local lockout, no key (the computer must either cancel local lockout, or
re-enable the display)
Factory Preset
and *RST: On
Turn the Full Screen Display On/Off
:DISPlay:MENU:STATe OFF|ON|0|1
:DISPlay:MENU:STATe?
Turns the full screen display mode on or off. Press
System, System to turn off full
screen mode.
History: Added with firmware revision A.08.00.