Agilent Technologies Model A.08.xx Water Dispenser User Manual


 
Chapter 5 281
Language Reference
[:SENSe]:DEMod Subsection
Demodulation Control
[:SENSe]:DEMod:STATe OFF|ON|0|1
[:SENSe]:DEMod:STATe?
Turns demodulation on or off.
Factory Preset
and *RST: Off
Front Panel
Access:
Det/Demod, Demod, Off
Demod Time
[:SENSe]:DEMod:TIME <time>
[:SENSe]:DEMod:TIME?
Sets the time used for frequency domain demodulation.
Factory Preset
and *RST: 500 ms
Range: 2 ms to 100 s
Default Unit: seconds
Front Panel
Access:
Det/Demod, Demod, Demod Time
Demod View
[:SENSe]:DEMod:VIEW[:STATe] OFF|ON|0|1
[:SENSe]:DEMod:VIEW[:STATe]?
This command causes the demodulated signal to be displayed. If FM Demod is on,
then the display scales the y-axis in units of kHz. The scale/div is set with the
command :DISPlay:WINDow:TRACe:Y
[:SCALe]:PDIVision:FREQuency <freq> if FM Demod is on. If FM
Demod is on, then several functions are not available; these include: Log/Lin
(display is always in linear), Y-Axis Units, Marker Search functions, Normalize,
Display Line, Peak Excursion, and Peak Threshold. There is no effect when AM
demodulation is used (only applicable for FM demodulation).
Factory Preset
and *RST: Off
Remarks: This command is not available when Demod is set to Off.
Front Panel
Access:
Det/Demod, Demod, FM, Demod View