[:SENSe]:AVERage Subsection
Turn Automatic Averaging On/Off
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Sets the averaging to be automatically set to the appropriate type for the current
measurement setup. Or allows you to manually choose the type of averaging with
[:SENSe]:AVERage:TYPE.
When AUTO is On:
If the Y Axis Scale is not Linear or Log, then average type is Video (Y Axis
Scale) Averaging.
If the Y Axis Scale is Linear or Log, then average type is Power Averaging.
If the Detector is Peak, Sample, or Negative Peak (not Average), then average
type is Video Average.
See Figure 5-2, which shows these auto rules for average type in flowchart format.
Figure 5-2 Auto Rules for Average Type