216 Chapter 5
Language Reference
CALCulate:MARKer Subsection
CALCulate:MARKer Subsection
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
Turns off all markers on all the traces.
Front Panel
Access:
Marker, Marker All Off
Continuous Peaking Marker Function
:CALCulate:MARKer[1]|2|3|4:CPEak[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:CPEak[:STATe]?
Turns on or off continuous peaking. It continuously puts the selected marker on the
highest displayed signal peak.
Factory Preset
and *RST: Off
Remarks: This command may not be used to activate a given marker.
Front Panel
Access:
Peak Search (or Search), Continuous Pk On Off
Frequency Counter Marker Resolution
:CALCulate:MARKer:FCOunt:RESolution <real>
:CALCulate:MARKer:FCOunt:RESolution?
Sets the resolution of the marker frequency counter. Setting the resolution to
AUTO will couple the marker counter resolution to the frequency span.
Factory Preset
and *RST: 1 kHz
Range: 1 Hz to 100 kHz
Default Unit: Hz
Front Panel
Access:
Freq Count, Resolution Auto Man