[:SENSe]:CORRection Subsection
Delete Amplitude Correction
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
Deletes the specified correction set. If the set was On, it is turned Off.
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections,
Antenna|Cable|Other|User, Delete Correction
Set Amplitude Correction Frequency Interpolation
[:SENSe]:CORRection:CSET[1]|2|3|4:X:SPACing
LINear|LOGarithmic
Sets the frequency interpolation to linear or logarithmic for the specified correction
set.
Remarks: Logarithmic frequency scale corrections are linearly
interpolated between correction points with respect to the
logarithm of the frequency. Linear frequency scale corrections
are interpolated along straight lines, connecting adjacent points
on a linear scale.
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Freq Interp Log Lin
Perform Amplitude Correction
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Turns the amplitude correction function on or off for the given set.
NOTE [:SENSe]:CORRection:CSET:ALL[:STATe] must be on for this command to
function.
Factory Preset
and *RST: Off
Remarks: CSET number equivalents to front panel access definitions are
as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections,
Antenna|Cable|Other|User, Correction On Off