STATus:QUEStionable Subsection
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition <integer>
:STATus:QUEStionable:CALibration:NTRansition?
This command determines which bits in the Questionable Calibration Condition
register will set the corresponding bit in the Questionable Calibration Event
register when that bit has a negative transition (1 to 0). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.
Factory Preset
and *RST: 0
Range: Integer, 0 to 32767
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition <integer>
:STATus:QUEStionable:CALibration:PTRansition?
This command determines which bits in the Questionable Calibration Condition
register will set the corresponding bit in the Questionable Calibration Event
register when that bit has a positive transition (0 to 1). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.
Factory Preset
and *RST: 32767 (all 1s)
Range: Integer, 0 to 32767
Questionable Condition
:STATus:QUEStionable:CONDition?
This query returns the decimal value of the sum of the bits in the Questionable
Condition register.
NOTE The data in this register is continuously updated and reflects the current conditions.