SYSTem Subsystem
Power On Elapsed Time
:SYSTem:PON:ETIMe?
Returns the number of seconds that have elapsed since the analyzer was turned on
for the very first time.
Front Panel
Access:
System, Show System
Power On Time
:SYSTem:PON:TIME?
Returns the number of milliseconds that have elapsed since the analyzer was last
turned on.
Power On Type
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Sets the defined instrument conditions after a power-on or
Preset.
PRESet - The instrument settings at power-on will be either the factory preset or
user preset, as set by :SYSTem:PRESet:TYPE FACTory|USER.
LAST - The instrument settings at power-on will be the settings at the time of
power down.
Factory Preset
and *RST: The factory default is Preset. This parameter is persistent, which
means that it retains the setting previously selected, even
through a power cycle.
Front Panel
Access:
System, Power On/Preset, Power On Last Preset
Enable IF/Video/Sweep Output Ports
:SYSTem:PORTs:IFVSweep:ENABle OFF|ON|0|1
:SYSTem:PORTs:IFVSweep:ENABle?
This command enables or disables the IF, video, and sweep output ports for
analyzers having options A4J (IF, Sweep, and Video Ports) and AYX (Fast Time
Domain Sweeps).
Factory Preset
and *RST: On
Example: :SYST:PORT:IFVS:ENAB ON
Range: On/Off