Agilent Technologies Model A.08.xx Water Dispenser User Manual


 
Chapter 5 241
Language Reference
DISPlay Subsystem
42.9 to 162 dBµV with zero reference level offset and max
mixer level = 97 dBµV.
Default Unit: Current active units
Remarks: The input attenuator setting may be affected. The minimum
displayed value of reference level is 327.6 dBm, and the
maximum displayed value is 327.6 dBm. See the remarks given
for the command
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
<rel_ampl>
Front Panel
Access:
Amplitude Y Scale, Ref Level
Trace Y-Axis Reference Level Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Sets the amplitude level offset for the y-axis.
Factory Preset
and *RST: 0 dB
Range: 327.6 to 327.6 dB
Default Unit: dB
Remarks: The sum of (reference level offset + reference level) is clipped to
the range 327.6 to 327.6 dB. The maximum limits are
determined by the setting of the first of these two parameters,
within the boundaries of their individual limits when initially
set.
For example, if the reference level is (first) set to 20 dBm, then
the reference level offset can be set to values of 307.6 dB to
327.6 dB. In the case of a 327.6 dB reference level offset, the
resultant reference level value changes to 307.6 dBm. The
reference level value range can be initially set to values from
149.9 to 55 dBm.
Setting the reference level offset value first yields the following:
If the reference level offset is (first) set to 30 dB, then the
reference level can be set to values of 327.6 to 25 dBm. The
reference level is clamped at 25 dBm because its positive
value of 55 dBm is reached at 25 dBm with an offset of 30 dB.
Its own positive amplitude limit applies.
If the reference level offset is (first) set to 30 dB, then the
reference level can be set to values of 327.6 to 85 dBm. Again,
the positive amplitude limit of reference level (alone) is factored
in to the resultant combined limit.