HP (Hewlett-Packard) 132A Water Dispenser User Manual


 
Chapter 3 Programming Your Universal Counter for Remote Operation
Elements of SCPI Commands
Programming Guide 3-35
Command Settings for Optimizing Throughput
This section lists the commands which enable the Counter to transfer data at the
fastest possible rate. See the To Optimize Throughputsample programs on
pages 3-73 , 3-86, and 3-96
Commands to Set Counter for Optimal Throughput
Unless otherwise noted, these settings are stored on Save (*SAV).
All of these settings are reset by *RST or a power cycle.
Disable auto trigger on measurement channel(s):
Specify absolute trigger levels
[:SENSe]:EVENt[1|2]:LEVel[:ABSolute] <numeric_value> [V]
or simply disable auto-trigger
[:SENSe]:EVENt[1|2]:LEVel[:ABSolute]:AUTO OFF
Set gate/arm to auto for appropriate measurement:
For Frequency, Period, and Ratio
[:SENSe]:FREQuency:ARM[:STARt]:SOURce IMMediate
[:SENSe]:FREQuency:ARM:STOP:SOURce IMMediate
For Phase
[:SENSe]:PHASe:ARM[:STARt]:SOURce IMMediate
For Time Interval (HP 53131A Only)
[:SENSe]:TINTerval:ARM[:STARt]:SOURce IMMediate
[:SENSe]:TINTerval:ARM:STOP:SOURce IMMediate
For Time Interval (HP 53132A Only)
[:SENSe]:TINTerval:ARM: ESTART:LAYer2:SOURce IMMediate
[:SENSe]:TINTerval:ARM :ESTART[:LAYer[1]]:SOURce IMMediate
[:SENSe]:TINTerval:ARM :ESTOP:LAYer2:SOURce IMMediate
[:SENSe]:TINTerval:ARM: :ESTOP[:LAYer[1]]:SOURce IMMediate
Define device trigger to FETC?:
When the device trigger is defined as FETC?, the Group Execute Trigger
should be used to query for a result.
*DDT #15FETC?